This spirometer is a compact devices based on different types of transducers (Fleisch or Turbine), a wide liquid crystal backlit screen and an internal or external printer. They can incorporate a module for measuring the Maximum inspiratory and expiratory pressures (MIP-MEP) and another optional module for pulse oximetry measurement (SpO2).
Optional software allows connection to a computer in real or deferred time, in order to perform and store the spirometric tests, or to transfer information by other means.
All the system is controlled by a microprocessor for the acquisition, calculation and display of alphanumeric and graphic data.
The system comes complete with transducer (Fleisch or Turbine type), 100 disposable mouthpieces, nose clip, 2 rolls of thermal paper
Technical Specifications :
Turbine type transducer, (Fleisch) type transducer is an optional
FVC, VC, MVV tests in PRE and POST bronchodilatation mode
Graphic superposition
Check-up program
Configuration program
Internal data base for 500 tests graphics (Option)
Internal printer : 58 mm
External printer connection (Option)
Data transfer via Internet with PC of : Patients tests, Device check-up data & Software upgrade
Tests exportation to other management systems
Calibration for quality control or tests, according to standards
SpO2 module (Option)
MEP-MIP module accessories (Option)
Software under Windows to work with PC in real or deferred time (Option)
Measuring Range : Flow from 0 to ± 16 l/s; volume from 0 to 10 l
Flow-Volume Accuracy : - 5% or 200ml/s - 3% or 50 ml - whichever greater (ATS, ERS)
Flow resistance : <0.06 kPa/l/s at 14 l/s (ATS, ERS)
Resistance at 14 l/s : - Turbine = 0.125 KPa/lls
Display : Liquid Crystal Display; 320x240 pixels, Area of 120x90 mm
Data input : Alphanumeric membrane keyboard
Maximum duration and number of manoeuvres : FVC, 25s (5 curves); VC, 45s (5 curves); MVV, 15s (5 curves)