MM-C002 Patient Monitor
Products Patient Monitor System MM-C002 Patient Monitor
|
| Features:
- Super thin (6cm), hanging wall design.
- Application of embedded system and high integrated intelligent modular technology.
- Patient information management function such as patient ID, age, sex, ward No, etc.
- Digital monitoring with powerful functions.
- Low power consumption and low malfunction rate.
- Automatic saving after sudden power-off.
- 12.1”or 15” high resolution TFT display with 9 waveforms.
- Selectable 9 modes of presentation, large font, graphic displaying of 7 leads in full screen.
- ST segment, 23 kinds of arrhythmia analysis, auto detection of lead drop.
- Memory and playback function of ECG waveforms.
- 720 hours of Trend graph display.
- Memory of 2880 groups of NIBP data.
- Function of Oxy-CRG analysis.
- Unique modes of direct, reverse or quick measuring, applicable for adult, infant and neonate.
- Function of overload protection.
- Anti high-frequency electro tome and defibrillate interference, can be used with pace-maker.
- Nurse-call function.
- Net-work function.
- Standard ECG, NIBP, SPO2, RESP, TEMP, PR.
- Option Belt-in thermal recorder, Dual-IBP, EtCo2.
Specification:
Size and Weight
- Size: 560*305*450mm
- Weight: 5.2kgd(net), 10.5(gross)
Recorder
- Built-in, thermal array
- Recording mode: manual, alarm and timing
- Parameter alarm recording
- Waveform freeze recording
- Trend list/graph recording
- Alarm event review recording
- NIBP review recording
- Monitor information recording
Display
- 12.1” or 15” TFT, 800*600
- Maximum 9 waveforms in full screen
- Power/standby indicator and battery indicator
Recall
- Trend recall: 720hours, I-minute resolution
- Alarm event recall: all parameters
- NIBP measurement recall: 1400 NIBP measurement data
Battery
- Rechargeable 12V Acid battery
- Regular working time above 2 hours
- Lasting 5 minutes after first alarm of low battery
- AC power supply: 85-260V 50-60Hz ±10%
Environment
- Operating Temperature 5~40°C Humidity = 80%
- Storage Temperature -20~55°C Humidity = 93%
ECG
- Displaying graphics of 7 leads on the same screen.
- Overload protection Defibrillation.
- Anti the interference of high-frequency electrotome
- Lead: 5 leads (RA, LA, V, RL, LL)
- Lead Selection: I; II; III; avR; avL; avF; V;
- Waveform:7 channels
- Gain Selection: ×0. 5; ×1; ×2; ×4
- Electrical Leakage: < 30uA
- Baseline Recovery Time: <3 sec following defibrillation.
- Protection: separates and withstands, 4000V AC/50Hz defibrillation, resistent to interference of electrotonus and myoelectricity.
- Scan Speed:12.5mm/25mm/50mm/sec
- Lead Drop: Auto detection
- Modes: DIA 0.05~100Hz,
MON 0.5~35Hz, OPE 1~25Hz
RESP
- Model: RA-LL impedance
- Measurement /alarm range:1—99 rpm
- Resolution:1rpm.
- Accuracy: ±2rpm.
NIBP
- Unique Direct, Reverse and Quick modes
- Method: automatic oscillometric
- Working Mode: Auto/Manual
- Interval of Auto Mode:1-60 min
- PR Range: 0-300 bpm
- Measurement Unit: mmHg/Kpa Selectable
- Measurement Ranges: Systolic BP,
- Diastolic BP and Mean BP
- Applicability: Adult/Pediatric/Neonatal
- Overload protection: Dual security protection
SPO2
- SPO2probe of high intensity and durability
- Ecellent measurement of low
perfusion.
- Anti interference of movement
- Measurement Range: 0-100%
- Alarm range: 0-100%
- Resolution: ±1%
- Accuracy: ±2% (80-100%);±3% (60-79%)
- PR range: 0~300bpm
HR
- Measurement Range: 0-300bpm
- Accuracy: ±1bpm or ±2%
- Resolution: 1bpm
- ECG Calibration: 1mV (Peak Value)
- ST Segment offset (Diagnose Mode)
- Measurement Range: -0.8mV~2.0mV
- Accuracy: ±0.02mV or ±5%
TEMP
- Resolution: 0.1 degrees Celsius
- Accuracy: ±0.2 degrees Celsius
- Channel: 2 channels
- Average measurement Time: < 10sec
- Alarm: Sound, light
|
| | |
|
|